ATOMS can perform several useful calculations using tables of x-ray absorption cross sections, including estimations of absorption length and various experimental corrections. There are several tables x-ray absorption cross sections which the user can choose from:
and the
technical documentation in Appendix
.
In an EXAFS experiment in transmission mode, proper sample preparation is essential for collection of high quality data. In order to make appropriate choices for the composition and form of the sample it is necessary to know both the absorption length of the sample and its edge step absorption length. ATOMS approximates both of these quantities.
The absorption length is defined as the thickness x of the
sample such that the intensity of the x-rays incident upon the sample
at an energy 50 eV above the absorption edge is attenuated e-fold.
That is,
where
is the total absorption of the
sample at that energy. The sample length for unit edge step, the
reciprocal of which is called
,
is the thickness x
such that there is an e-fold change in absorption between 50 eV below
and above the edge, i.e. for
.
ATOMS computes the free atom cross-sections for each atom specified in the input data at 50 eV above the absorption edge and for the resonant atom at 50 eV below the edge. Using the crystallographic information, it knows the size of the unit cell and how many of each species are in the unit cell and so can correctly calculate the total and unit edge step absorption lengths for the crystal. Using the unit cell size and tabulated masses of the elements, the density of the crystal is also computed. With the absorption lengths and density of the material, proper choices for sample preparation can be made.
Here is an example of a sample of lead titanate, PbTiO3, prepared using these calculations from ATOMS. Using this structural data
title PbTiO3 25C
title Glazer and Mabud, Acta Cryst. B34, 1065-1070 (1978)
core=ti Space P 4 m m
a=3.905 nitrogen=1 rmax=3.6 c=4.156
atom
! At.type x y z tag
Pb 0.0 0.0 0.0
Ti 0.5 0.5 0.539
O 0.5 0.5 0.1138 axial
O 0.0 0.5 0.6169 planar
yields the results shown in Table
.
| calculation | value |
| total |
4850.42 cm-1 |
| 747.42 cm-1 | |
| specific gravity | 7.942 |
| Normalization correction | 0.00046 Å2 |
| I0 correction | 0.00111 Å2 |
| self absorption correction | 0.00003 Å2 |
| self absorption amplitude factor | 1.093 |
| net correction | 0.00160 Å2 |
In this example, the absorption length is 1/4850 cm or about 2.1
microns and the edge step absorption length is 1/747 cm or about 13.4
microns. For my experiment on the titanium K edge of lead
titanate, I chose to make a sample which was 5 microns thick, thus
with total absorption of about 2.5 and an edge step of about 0.37.
Knowing the appropriate thickness for the sample, the dimensions of
the die used to press the sample, and the density of lead titanate, I
was able to make an sample appropriate to these conditions. An
example spectrum obtained with one of these samples is shown in
Fig.
, which has an edge step calculated by the
AUTOBK program of 0.413, only 10% larger than the
estimation made by ATOMS.
To avoid introducing systematic errors into the amplitude of the
measured
,
an edge-step normalization is typically used.
Since the true atomic background absorption,
,
has energy
dependence, normalization by the edge-step introduces an
energy-dependent attenuation to the amplitude of
.
This
attenuation is small for heavy elements, but can be of the same order
as thermal effects for light elements. ATOMS calculates an
approximation to this attenuation called the normalization
correction11.2.
Using the tables of free atoms x-ray cross-sections, ATOMS
evaluates the cross-section for the free central atom in a range from
50 to 300 eV above the absorption edge. It then regresses a second
order polynomial in the natural log of the energy relative to the edge
to the natural log of the free atom cross-section11.3. The linear
term is an approximation to the degree of attenuation introduced by
the edge-step normalization. This is intended as additive corrections
to the measured
of a fit. Neglecting this correction will
make the temperature dependence of
deviate from an Einstein
behavior by a constant offset.
In a fluorescence experiment the absorption cross section is obtained
from the detected intensities on the I0 and IF chambers by
measuring their ratio as a function of energy. This introduces an
energy dependent error, usually an attenuation, into the amplitude of
the measured
.
The secondary photon measured in the
fluorescence experiment is always of the same energy. Thus there is
no energy dependent part of the signal measured on IF. There is,
though, an energy response to I0 that is neglected when IF is
normalized by the signal on I0. ATOMS calculates an
approximation to this attenuation called the I0 correction.
To enable this calculation when using ATOMS, it is necessary to specify the contents of the I0 chamber. In ATOMS and TkATOMS this is enabled by specifying non-zero values for one or more of the fill gasses. From the supplied values of fill gases, ATOMS approximates the energy response of the I0 chamber using the tables of free-atom x-ray cross sections.
ATOMS evaluates the cross-section for the gases in the I0chamber in a range from 50 to 300 eV above the absorption edge. It
then regresses a second order polynomial in the natural log of the
energy relative to the edge to the natural log of the gas
cross-section. The linear term is an approximation to the degree of
attenuation introduced by normalizing IF by I0. This is
intended as an additive correction to the measured
of a
fit. Neglecting this correction will make the temperature dependence
of
deviate from an Einstein behavior by a constant offset.
The second correction required for fluorescence measurements is called
the self-absorption correction. This calculation is also
enabled when fill gases are specified. Assuming11.4 equal entry and
exit angles, the formula for the fluorescence signal is given by
is the absorption of the resonant atom,
is the
rest of the absorption in the material, and
is the absorption
of the material at the fluorescence energy of the resonant atom. The
self-absorption correction is due to the
term in the
denominator of this equation. In the limit that
,
the self-absorption correction is
enormous, canceling the oscillatory structure of the XAFS spectrum.
ATOMS approximates the effect of
term in the denominator of
Eq. (
) given the assumption of infinite sample
thickness, which, in practice, means that the sample is very thick
compared to the absorption length.
The correction presented in this section is a correction to the
measured
.
A similar correct for the XANES portion of the
spectrum can be calculated, but that is not the purpose of
this one. The signal
is the normalized variation in
IF/I0, the oscillatory part of the absorption of the resonant
atom. I will now derive expressions for the variation in
IF/I0 and for the normalization term using the notation of
Eq. (
).
The variation in the signal,
is due to the
variation in
,
In this equation
.
The normalization,
,
is the difference in
IF/I0 below
and above the edge. The symbols
and
denote that
the quantity is evaluated above or below the absorption edge.
), but
| (11.3) |
Finally I write an expression for the normalized variation in
IF/I0 using Eqs. (
) and
(
). Since this derivation is for a correction to
,
I am only concerned with normalized variation above the
absorption edge.
The ideal measurement would be undistorted and simply expressed as
.
The correction factor
is the factor by which the measured signal
must be multiplied to obtain the ideal signal.
ATOMS uses the free atom cross-sections to evaluate
Eq. (
) in the energy range from 50 to 300 eV
above the absorption edge11.5. A second order polynomial in the natural log of the
energy relative to the edge is regressed to the natural log of
Eq. (
).
The amplitude factor is the exponent of the constant term in the regression. This is intended as a multiplicative correction to the amplitude of the data. If this term is neglected in a fit, the measured S02 will be too small by that factor. If the density of the absorbing atom is large, this factor can be huge. Indeed for pure transition metal K edges, the oscillatopry structure is damped out almost completely.
The other correction is the linear term of the regression and is
expressed as a second order cumulants. This is intended as an
additive corrections to the measured
of a fit. Neglecting
this correction will make the temperature dependence of
deviate from an Einstein behavior by a constant offset.